New users
Please contact Maya Holdengreber, tel. 073-378-1107 to coordinate a meeting.
Please do not make a reservation through our online BookItLab system.
The Olympus Slideview VS200 is a slide scanner for high-throughput digitalization of histology and fluorescence slides for archival, diagnostics, teaching, and research. Our system supports three imaging modes: Brightfield, Polarization, and Fluorescence.
The system’s automatic slide loader can hold standard slides (25-26.5 mm × 75-76.5 mm) and also 51-53 mm x 75-76.5 mm double slides. High-resolution images are converted into virtual slides through an automated high-speed scanning and tiling process.
This is a proprietary system.
Excelitas X-Cite NOVEM 9-Channel LED Illumination System.
LED | Emission filter | Fluorophore family |
385 | 432/36 | DAPI |
475 | 515/30 | FITC |
555 | 595/31 | Cy3, mCherry |
635 | 680/42 | Cy5 |
735 | 809/81 | Cy7 |
Objective | Magnification/NA | Working Distance | Pixel size Brightfield | Pixel size Fluorescence |
---|---|---|---|---|
Universal Plan Extended Super Apochromat | x10/0.4 | 3.1 mm | 548 nm | 548 nm |
Universal Plan Extended Apochromat | x20/0.80 | 0.6 mm | 273.8 nm | 325 nm |
Universal Plan Extended Super Apochromat | x40/0.45 | 0.18 mm | 137 nm | 172.5 nm |
Camera | Technology | Application | Resolution | Data depth |
---|---|---|---|---|
iDS VS-264C | CMOS | Brightfield | ||
CAM-ORCA-FusionBT Digital Hamamatsu | sCMOS | Fluorescence | 2304 x 2304 px | 16 bit |
All fixed specimen brightfield/widefield applications including FISH and rare events, polarized light, phase contrast, darkfield. One plane, z-stack or extended focus acquisition.
Acquisition is performed using the proprietary CellSens software. The .vsi virtual slide image files can be opened by Olyvia, QuPath, FIJI software.
Please contact Maya Holdengreber, tel. 073-378-1107 to coordinate a meeting.